CVE-2019-25722 | BDVDB-2026-337158 Detail
DrägerSC 6002XL
Description#
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.
Impact#
Per CVSS v3.1: confidentiality impact low, integrity impact low, availability impact high. Attack vector: adjacent.
Weaknesses#
- CWE-798 Use of Hard-coded Credentials
References#
Exploit#
No public exploit known
History#
- Version 1Record published in the CVE List.
- Version 2Most recent update recorded in the CVE List.