CVE-2019-25721 | BDVDB-2026-337157 Detail
DrägerInfinity M300
Description#
Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.
Impact#
Per CVSS v3.1: confidentiality impact none, integrity impact none, availability impact high. Attack vector: adjacent.
Weaknesses#
- CWE-400 Uncontrolled Resource Consumption
References#
Exploit#
No public exploit known
History#
- Version 1Record published in the CVE List.
- Version 2Most recent update recorded in the CVE List.