This site currently shows sample data for demonstration.

Vulnerabilities

CVE-2019-25721 | BDVDB-2026-337157 Detail

DrägerInfinity M300

Description#

Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.

Impact#

Per CVSS v3.1: confidentiality impact none, integrity impact none, availability impact high. Attack vector: adjacent.

Weaknesses#

  • CWE-400 Uncontrolled Resource Consumption

References#

Exploit#

No public exploit known

History#

  1. Version 1Record published in the CVE List.
  2. Version 2Most recent update recorded in the CVE List.