CVE-2014-5406 | BDVDB-2026-70274 Detail
HospiraLifeCare PCA Infusion System
Description#
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Impact#
No impact statement in the source data.
Weaknesses#
References#
- https://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm
- https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.json
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01
Exploit#
No public exploit known
History#
- Version 1Record published in the CVE List.
- Version 2Most recent update recorded in the CVE List.