CVE-2019-25720 | BDVDB-2026-337379 Detail
DrägerSC 6002XL
Description#
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.
Impact#
Per CVSS v3.1: confidentiality impact none, integrity impact none, availability impact high. Attack vector: adjacent.
Weaknesses#
- CWE-1286 Improper Validation of Syntactic Correctness of Input
References#
Exploit#
No public exploit known
History#
- Version 1Record published in the CVE List.