CVE-2019-25724 | BDVDB-2026-337160 Detail
DrägerInfinity M300
Description#
Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.
Impact#
Per CVSS v3.1: confidentiality impact none, integrity impact none, availability impact high. Attack vector: adjacent.
Weaknesses#
- CWE-400 Uncontrolled Resource Consumption
References#
Exploit#
No public exploit known
History#
- Version 1Record published in the CVE List.
- Version 2Most recent update recorded in the CVE List.