CVE-2025-14096 | BDVDB-2026-306502 Detail
Radiometer Medical ApsABL90 FLEX and ABL90 FLEX PLUS Analyzers
Description#
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Impact#
CAPEC-69 — Target Programs with Elevated Privileges
Weaknesses#
References#
Exploit#
No public exploit known
History#
- Version 1Record published in the CVE List.