This site currently shows sample data for demonstration.

Vulnerabilities

CVE-2025-1204 | BDVDB-2026-267421 Detail

Contec HealthCMS8000 Patient Monitor

Description#

The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.

Impact#

Per CVSS v4.0: confidentiality impact high, integrity impact high, availability impact high. Attack vector: network.

Weaknesses#

References#

Exploit#

No public exploit known

History#

  1. Version 1Record published in the CVE List.