CVE-2023-0888 | BDVDB-2026-197220 Detail
B. BraunSpace Battery Pack SP with Wi-Fi
Description#
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
Impact#
CAPEC-233 Privilege Escalation
Weaknesses#
- CWE-95 Eval injection
References#
Exploit#
No public exploit known
History#
- Version 1Record published in the CVE List.