CVE-2015-3956 | BDVDB-2026-114113 Detail
HospiraPlum A+ Infusion System
Description#
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Impact#
No impact statement in the source data.
Weaknesses#
- CWE-345 Insufficient verification of data authenticity CWE-345
References#
Exploit#
No public exploit known
History#
- Version 1Record published in the CVE List.
- Version 2Most recent update recorded in the CVE List.